array(2) { ["lab"]=> string(4) "1409" ["publication"]=> string(5) "12599" } Determination of short-chain chlorinated paraffins in multiple matrices of Arctic using gas chromatography-electron capture negative ion-low resolution mass spectrometry - Liang Yong | LabXing

Determination of short-chain chlorinated paraffins in multiple matrices of Arctic using gas chromatography-electron capture negative ion-low resolution mass spectrometry

2018
期刊 MethodsX
Gas chromatography-electron capture negative ion-low resolution mass spectrometry (GC-ECNI-MS) was used for the quantification of short-chain chlorinated paraffins (SCCPs) in multiple matrices of Arctic. Samples were spiked with surrogate standards (13C10-trans-chlordane) and extracted with dichloromethane and hexane (1:1, v/v) using accelerated solvent extraction (ASE). The extract was cleaned using a multilayer Silica-Florisil column and e-HCH was added before instrument analysis. The SCCPs were analyzed using a gas chromatograph (GC) in electron capture negative ion (ECNI) mode coupled with a 7000B triple quadruple mass spectrometer (MS) in single quad mode. The calibration is performed using three commercial standards (chlorine contents of 51.5%, 55.5%, and 63.0%). A reasonable linear correlation for commercial standards was found between chlorine content and total response factor …

  • 卷 5
  • 页码 939-943
  • Elsevier